Online Gas Analyzer

Manufacturer: AP2E
AddThis Social Bookmark Button
DetailsCataloge
SAMPLING
Flow Rate 3-9 L/h
Max. Temperature 600°C
Max. Humidity H2O(g) < 40% vol. - Standard
H2O(g) < 40% vol. - Study Required
Pressure 1 atm. ± 100 mbar @ sampling point
Sampling Line Ambient Temp. > 10°C et H2O < 40% vol.
> Simple polytube (no heating)
Ambient Temp. < 10°C et H2O > 40% vol.
> 40°C heated line
DIMENSIONS
Size Standard 19”, 4U rack.
550mm depth.
Weight 20kg
Options Wall mounted
ATEX compliant integration
ELECTRONICS
Display/Control 5.7” diagonal color touch screen
PC OS Windows® XP®
Software WinProceas ©
INSTALLATION REQUIREMENTS
Operating Temp. 15-35°C - Standard
10-40°C - Optional
Power Supply 200W - 110-220VAC - 50-60Hz
Compressed Air 1-6 bar (oil free). Not provided.
I/O's
Standard Ethernet Protocol; RS485
ModBus; RS232
Optional Analog I/O; TDR I/O
Other I/O’s on request
ANALYTICAL SPECIFICATIONS
Gas Rangea LODb
Formaldehyde 0 to 10 ppm 2 ppb
H2S 0 to 50 ppm 2 ppb
CH4 0 to 50 ppm 1 ppb
CO 0 to 100 ppm 1 ppb
CO2 0 to 1% vol 5 ppb
H2 0 to 5% vol 3 ppb
H2O 0 to 1000 ppm 5 ppb
HCI 0 to 5 ppm 1 ppb
HCN 0 to 10 ppm 1 ppb
HF 0 to 10 ppm 1 ppb
N2O 0 to 50 ppm 2 ppb
NH3 0 to 50 ppm 1 ppb
O2 0 to 1000 ppm 5 ppb
NO 0 to 5000 ppm 100 ppb
NO2 0 to 100 ppm 10 ppb